2022 Good Quality Reference Thermometers - PR1231/PR1232 Standard Platinum-10% Rhodium/Platium Thermocouple – Panran
2022 Good Quality Reference Thermometers - PR1231/PR1232 Standard Platinum-10% Rhodium/Platium Thermocouple – Panran Detail:
PR1231/PR1232Standard Platinum-10% Rhodium/Platium Thermocouple
Part1 Overview
The first and second grade standard platinum-iridium 10-platinum thermocouples which have high accuracy good physical and chemical properties, good oxidation resistance at high temperatures, good stability and reproducibility of thermoelectromotive force. Therefore, it is used as a standard measuring instrument in (419.527~1084.62) °C , it is also used to temperature magnitude transmit and precision temperature measure in the temperature range.
Part2 Technical parameters
|
Parameter index |
First grade platinum-iridium 10-platinum thermocouples |
Second grade platinum-iridium 10-platinum thermocouples |
|
Positive and negative |
The positive is a platinum-rhodium alloy (platinum 90% rhodium 10%), the negative is pure platinum |
|
|
electrode |
Two electrodes diameter is 0.5-0.015mm length is no less than 1000mm |
|
|
Requirements of the thermal electromotive force Measure junction temperature is at Cu point(1084.62℃)Al point(660.323℃) Zn point(419.527℃) and reference junction temperature is 0℃ |
E(tCu)=10.575±0.015mV E(tAl)=5.860+0.37 [E(tCu)-10.575]±0.005mV E(tZn)=3.447+0.18 [E(tCu)-10.575]±0.005mV |
|
|
Stability of Thermo-electromotive force |
3μV |
5μV |
|
Annual change Thermo-electromotive force at Cu point(1084.62℃) |
≦5μV |
≦10μV |
|
Working Temperature Range |
300~1100℃ |
|
|
Insulating sleeve |
Double hole porcelain tube or corundum tube Outer diameter(3~4)mm, hole diameter(0.8~1.0)mm, length(500~550)mm |
|
Part3 Application Instructions
standard platinum-iridium 10-platinum thermocouples must be in accordance with the national delivery system table,National verification procedures must be implemented. First grade standard platinum-iridium 10-platinum thermocouples can be used to measure second grade,Ⅰ grade,Ⅱ grade platinum-iridium 10-platinum thermocouples and Ⅰ grade base metal thermocouples; Second grade platinum-iridium 10-platinum thermocouples can only used to measure Ⅱ grade base metal thermocouples
|
National verification code |
National verification name |
|
JJG75-1995 |
standard platinum-iridium 10-platinum thermocouples calibration specification |
|
JJG141-2013 |
Working precious metal thermocouples calibration specification |
|
JJF1637-2017 |
base metal thermocouple calibration specification |
Part4 Maintenance and preservation
1. The standard thermocouple calibration period is 1 year, and every year the standard thermocouple must be calibrated by metrology department.
2. The necessary supervisory calibration should be performed according to the usage.
3. The work environment of the standard thermocouple should be clean to avoid contamination of the standard thermocouple.
4. The standard thermocouple should be placed in a non-polluting condition and protected from mechanical stress.
Part5 Precautions when use
1. The insulation tube cannot be used at high temperature roasting. The original insulation tube is used after strict cleaning and high temperature roasting.
2. The insulation tube ignores positive and negative, which will cause the platinum pole to be contaminated and the thermoelectric potential value to decrease.
3. Randomly the standard thermocouple insulation tube with cheap wire will contaminate the standard thermocouple, and the protective metal tube must be used for the verification of the base metal thermocouple.
4. The standard thermocouple cannot be suddenly placed in the temperature-regulating furnace, nor be taken out from the temperature-regulating furnace. The sudden-heat and cold will affect the thermoelectric performance
5.Under normal circumstances, the verification furnace for the precious metal thermocouple and the base metal thermocouple should be strictly distinguished; if it is impossible, the clean ceramic tube or corundum tube(diameter about 15mm) should be inserted to the furnace tube for protecting precious metal Thermocouples and standard thermocouples from base metal thermocouple pollution.
Product detail pictures:
Related Product Guide:
Dedicated to strict top quality command and considerate purchaser support, our experienced staff customers are always available to discuss your necessities and be certain full client gratification for 2022 Good Quality Reference Thermometers - PR1231/PR1232 Standard Platinum-10% Rhodium/Platium Thermocouple – Panran , The product will supply to all over the world, such as: Argentina, belarus, Buenos Aires, Qualified R&D engineer will be there for your consultation service and we will try our best to meet your requirements. So please feel free to contact us for inquiries. You'll be able to send us emails or call us for small business. Also you are able to come to our business by yourself to get further knowing of us. And we will surely give you the best quotation and after-sale service. We're ready to build stable and friendly relations with our merchants. To achieve mutual success, we will make our best efforts to build a solid co-operation and transparent communication work with our companions. Above all, we are here to welcome your inquiries for any of our goods and service.