We now have a highly efficient crew to deal with inquiries from clients. Our intention is 100% shopper pleasure by our merchandise quality, price tag & our staff service and take pleasure in a very good standing amongst purchasers. With quite a few factories, we can easily provide a wide vary of Three-Temperature-Zone Calibration Furnace, Refrigerated Temperature Calibration Bath, Calibrator Pressure Gauge, If you have the requirement for virtually any of our items, make sure you call us now. We are wanting ahead to hearing from you before long.
Good quality Reference Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran Detail:

Standard Platinum Resistance Thermometer

I. Description

The Standard Platinum Resistance Thermometer is used for compensation in standard temperature range of 13.8033k—961.8°C, and used as a standard when testing a variety of standard thermometers and high-precision thermometers. Within the above the temperature zone, it is also used directly for measuring the temperature of high-accuracy.

The Standard Platinum Resistance Thermometer measures the temperature according to the changeable regularity of the resistance temperature of the platinum.

In accordance with the regulations of ITS90, the T90 is defined by the platinum thermometer when triple point (13.8033K) of the nitrogen balance reaches to the temperature range of silver freezing point. It is indexed by using the group of required defined freezing point and reference function as well as the deviation function of the temperature interpolation.

The above temperature zoning is divided into several and decided to work normally within the sub-temperature zone by various different forms of structure of thermometers.

See the detailed thermometers in the table below:

Type

Classification

Suitable Temperature Zone

Working Length

(mm)

Temperature

WZPB-1

I

0~419.527

470±10

Medium

WZPB-1

I

83.8058K~419.527

470±10

Full

WZPB-2

II

0~419.527

470±10

Medium

WZPB-2

II

83.8058K~419.527

470±10

Full

WZPB-7

I

0~660.323

510±10

Medium

WZPB-7

I

83.8058K~660.323

510±10

Full

WZPB-8

II

0~660.323

510±10

Medium

WZPB-8

II

83.8058K~660.323

510±10

Full

Note: The Rtp of the above thermometers is 25±1.0Ω.The external diameter of quartz tubes is φ7±0.6mm.our factory also manufactures the platinum thermometer with the temperature zone of 83.8058K~660.323 as the working basic standard instrument.

II. Use Information

1. Before using, firstly, check the thermometer number to be consistent with the test certificate.

2. When using, according to the lug logo of the thermometer wire terminal, connect the wire correctly. The lug① of the red wire is connected to the current positive terminal; the lug of the yellow wire, to the current negative terminal; and the lug of the black wire, to the potential positive terminal; the lug of the green wire, to the potential negative terminal.

The following is the outline of the thermometer:

SPRT basic structure.png

3. The current should be 1MA according to the measuring of the temperature component of the thermometer.

4. For matching the electrical measurement device of the thermometer for measuring temperature, the low resistance potentiometer of grade 1 and the standard coil resistance of grade 0.1 or the measuring precise temperature bridge as well as the accessories should be used. The complete set of electrical measuring device should be guaranteed to have the sensitivity to distinguish the changing of one ten-thousandth Ohm.

5. During the process of using, preservation and transportation, try to avoid the severe mechanical vibration of the thermometer.

6. When using the first grade Standard Platinum Resistance Thermometer to test the temperature of the second grade Standard Platinum Resistance Thermometer, should follow the verification procedures approved by National Measurement Bureau.

7. The regular test of the thermometer should be done stringently according to the relevant verification procedures and regulations .


Product detail pictures:

Good quality	Reference Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Good quality	Reference Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Good quality	Reference Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Good quality	Reference Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Good quality	Reference Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures


Related Product Guide:

We have now our possess revenue group, design staff, technical crew, QC team and package group. We now have strict excellent regulate procedures for each process. Also, all of our workers are experienced in printing subject for Good quality Reference Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran , The product will supply to all over the world, such as: The Swiss, Tunisia, Adelaide, In order to meet our market demands, we have paied more attention to the quality of our products and services. Now we can meet customers' special requirements for special designs. We persistently develop our enterprise spirit quality lives the enterprise, credit assures cooperation and keep the motto in our minds: customers first.
With a positive attitude of regard the market, regard the custom, regard the science, the company works actively to do research and development. Hope we have a future business relationships and achieving mutual success.
5 Stars By Cindy from Israel - 2018.02.12 14:52
Customer service staff and sales man are very patience and they all good at English, product's arrival is also very timely, a good supplier.
5 Stars By ROGER Rivkin from Montpellier - 2017.03.28 16:34