Model |
PR9112 Intelligent Pressure Calibrator |
|
|
Pressure Measurement |
Measurement Range |
(-0.1~250)Mpa |
|
Display accuracy |
±0.05%F.S, ±0.02%F.S |
|
|
Electric Current Measurement |
Range |
±30.0000mA |
|
Sensitivity |
0.1uA |
|
|
Accuracy |
±(0.01%R.D+0.003%F.S) |
|
|
Voltage Measurement |
Range |
±30.0000V |
|
Sensitivity |
0.1mV |
|
|
Accuracy |
±(0.01%R.D +0.003%F.S) |
|
|
Switching Value |
A group of Power/outage measurement |
|
|
Output function |
Direct-current output |
DC24V±0.5V |
|
Operating Environmental |
Working Temperature |
(-20~50)℃ |
|
Relative Temperature |
<95% |
|
|
Storage Temperature |
(-30~80)℃ |
|
|
Power Supply Configuration |
Power supply mode |
Lithium battery or power supply |
|
Battery Operating Time |
60 hours (24V with no load) |
|
|
Charging time |
About 4 hours |
|
|
Other indicators |
Size |
115mm×45mm×180mm |
|
Communication interface |
specialized three-core aviation plug |
|
|
Weight |
0.8KG |
|
Main Application:
1.Calibrate pressure (differential pressure) transmitter
2.Calibrate pressure switch
3.Verify the precision pressure gauge, General pressure gauge.
Product Feature:
1.Built-in manual operator function, The HART intelligent pressure transmitter can be calibrated.(optional)
2.Double-row liquid crystal display with backlight.
3.mmH2O、mmHg、psi、kPa、MPa、Pa、mbar、bar、kgf/c㎡, switch between nine pressure units.
4.With DC24V output function.
5.With current, voltage measurement.
6.Measuring with switching volume.
7.With communication interface.(optional)
8. Storage capacity: total 30 pcs file, (50 data records of each file)
9.Big screen crystal liquid display
Software Configuration:
PR9112S software of pressure verification system is the supporting software of our digital pressure calibrator series products in our company, Data collection records can be performed, Auto generated form, Automatic error calculation, Print certificate.
1.Routine Pressure Range Selection Table
|
No. |
Pressure Range |
Type |
Class of accuracy |
|
01 |
(-100~0) kPa |
G |
0.02/0.05 |
|
02 |
(0~60)Pa |
G |
0.2/0.05 |
|
03 |
(0~250)Pa |
G |
0.2/0.05 |
|
04 |
(0 ~ 1) kPa |
G |
0.05/0.1 |
|
05 |
(0 ~ 2) kPa |
G |
0.05/0.1 |
|
06 |
(0 ~ 2.5) kPa |
G |
0.05/0.1 |
|
07 |
(0 ~ 5) kPa |
G |
0.05/0.1 |
|
08 |
(0 ~ 10) kPa |
G |
0.05/0.1 |
|
09 |
(0 ~ 16) kPa |
G |
0.05/0.1 |
|
10 |
(0 ~ 25) kPa |
G |
0.05/0.1 |
|
11 |
(0 ~ 40) kPa |
G |
0.05/0.1 |
|
12 |
(0 ~ 60) kPa |
G |
0.05/0.1 |
|
13 |
(0 ~ 100) kPa |
G |
0.05/0.1 |
|
14 |
(0 ~ 160) kPa |
G/L |
0.02/0.05 |
|
15 |
(0 ~ 250) kPa |
G/L |
0.02/0.05 |
|
16 |
(0 ~ 400) kPa |
G/L |
0.02/0.05 |
|
17 |
(0 ~ 600) kPa |
G/L |
0.02/0.05 |
|
18 |
(0 ~ 1) MPa |
G/L |
0.02/0.05 |
|
19 |
(0 ~ 1.6) MPa |
G/L |
0.02/0.05 |
|
20 |
(0 ~ 2.5) MPa |
G/L |
0.02/0.05 |
|
21 |
(0 ~ 4) MPa |
G/L |
0.02/0.05 |
|
22 |
(0 ~ 6) MPa |
G/L |
0.02/0.05 |
|
23 |
(0 ~ 10) MPa |
G/L |
0.02/0.05 |
|
24 |
(0 ~ 16) MPa |
G/L |
0.02/0.05 |
|
25 |
(0 ~ 25) MPa |
G/L |
0.02/0.05 |
|
26 |
(0 ~ 40) MPa |
G/L |
0.02/0.05 |
|
27 |
(0 ~ 60) MPa |
G/L |
0.05/0.1 |
|
28 |
(0 ~ 100) MPa |
G/L |
0.05/0.1 |
|
29 |
(0 ~ 160) MPa |
G/L |
0.05/0.1 |
|
30 |
(0 ~ 250) MPa |
G/L |
0.05/0.1 |
Remarks: G=GasL=Liquid
2.Composite Pressure Range Selection Table:
|
No. |
Pressure Range |
Type |
Class of accuracy |
|
01 |
±60 Pa |
G |
0.2/0.5 |
|
02 |
±160 Pa |
G |
0.2/0.5 |
|
03 |
±250 Pa |
G |
0.2/0.5 |
|
04 |
±500 Pa |
G |
0.2/0.5 |
|
05 |
±1kPa |
G |
0.05/0.1 |
|
06 |
±2kPa |
G |
0.05/0.1 |
|
07 |
±2.5 kPa |
G |
0.05/0.1 |
|
08 |
±5kPa |
G |
0.05/0.1 |
|
09 |
±10kPa |
G |
0.05/0.1 |
|
10 |
±16kPa |
G |
0.05/0.1 |
|
11 |
±25kPa |
G |
0.05/0.1 |
|
12 |
±40kPa |
G |
0.05/0.1 |
|
13 |
±60kPa |
G |
0.05/0.1 |
|
14 |
±100kPa |
G |
0.02/0.05 |
|
15 |
(-100 ~160) kPa |
G/L |
0.02/0.05 |
|
16 |
(-100 ~250) kPa |
G/L |
0.02/0.05 |
|
17 |
(-100 ~400) kPa |
G/L |
0.02/0.05 |
|
18 |
(-100 ~600) kPa |
G/L |
0.02/0.05 |
|
19 |
(-0.1~1)Mpa |
G/L |
0.02/0.05 |
|
20 |
(-0.1~1.6)Mpa |
G/L |
0.02/0.05 |
|
21 |
(-0.1~2.5)Mpa |
G/L |
0.02/0.05 |
Remarks:
1.Partial range can do absolutely pressure
2.Automatic temperature compensation range:(-20~50℃)
3.Pressure transfer medium requires non-corrosive
<!–
Packing
Product detail pictures:
Related Product Guide:
We now have a skilled, performance team to supply good quality services for our consumer. We often follow the tenet of customer-oriented, details-focused for OEM/ODM China High Accuracy Digital Pressure Gauge - PR9112 Intelligent Pressure Calibrator – Panran , The product will supply to all over the world, such as: Netherlands, Washington, Portland, With the goal of zero defect. To care for the environment, and social returns, care employee social responsibility as own duty. We welcome friends from all over the world to visit and guide us so that we can achieve the win-win goal together.