Good quality To start with,and Purchaser Supreme is our guideline to offer the top service to our customers.Presently, we've been seeking our best to be amongst the top exporters inside our industry to fulfill consumers extra need to have for Temperature Instrument Calibrator, Temperature Calibrators & Furnaces, Pressure Pump Calibrator, We warmly welcome domestic and abroad buyers deliver inquiry to us ,we now have 24hours doing work team! Anytime anywhere we are still here to be your partner.
Wholesale Price Ir Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran Detail:

Standard Platinum Resistance Thermometer

I. Description

The Standard Platinum Resistance Thermometer is used for compensation in standard temperature range of 13.8033k—961.8°C, and used as a standard when testing a variety of standard thermometers and high-precision thermometers. Within the above the temperature zone, it is also used directly for measuring the temperature of high-accuracy.

The Standard Platinum Resistance Thermometer measures the temperature according to the changeable regularity of the resistance temperature of the platinum.

In accordance with the regulations of ITS90, the T90 is defined by the platinum thermometer when triple point (13.8033K) of the nitrogen balance reaches to the temperature range of silver freezing point. It is indexed by using the group of required defined freezing point and reference function as well as the deviation function of the temperature interpolation.

The above temperature zoning is divided into several and decided to work normally within the sub-temperature zone by various different forms of structure of thermometers.

See the detailed thermometers in the table below:

Type

Classification

Suitable Temperature Zone

Working Length

(mm)

Temperature

WZPB-1

I

0~419.527

470±10

Medium

WZPB-1

I

83.8058K~419.527

470±10

Full

WZPB-2

II

0~419.527

470±10

Medium

WZPB-2

II

83.8058K~419.527

470±10

Full

WZPB-7

I

0~660.323

510±10

Medium

WZPB-7

I

83.8058K~660.323

510±10

Full

WZPB-8

II

0~660.323

510±10

Medium

WZPB-8

II

83.8058K~660.323

510±10

Full

Note: The Rtp of the above thermometers is 25±1.0Ω.The external diameter of quartz tubes is φ7±0.6mm.our factory also manufactures the platinum thermometer with the temperature zone of 83.8058K~660.323 as the working basic standard instrument.

II. Use Information

1. Before using, firstly, check the thermometer number to be consistent with the test certificate.

2. When using, according to the lug logo of the thermometer wire terminal, connect the wire correctly. The lug① of the red wire is connected to the current positive terminal; the lug of the yellow wire, to the current negative terminal; and the lug of the black wire, to the potential positive terminal; the lug of the green wire, to the potential negative terminal.

The following is the outline of the thermometer:

SPRT basic structure.png

3. The current should be 1MA according to the measuring of the temperature component of the thermometer.

4. For matching the electrical measurement device of the thermometer for measuring temperature, the low resistance potentiometer of grade 1 and the standard coil resistance of grade 0.1 or the measuring precise temperature bridge as well as the accessories should be used. The complete set of electrical measuring device should be guaranteed to have the sensitivity to distinguish the changing of one ten-thousandth Ohm.

5. During the process of using, preservation and transportation, try to avoid the severe mechanical vibration of the thermometer.

6. When using the first grade Standard Platinum Resistance Thermometer to test the temperature of the second grade Standard Platinum Resistance Thermometer, should follow the verification procedures approved by National Measurement Bureau.

7. The regular test of the thermometer should be done stringently according to the relevant verification procedures and regulations .


Product detail pictures:

Wholesale Price	Ir Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale Price	Ir Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale Price	Ir Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale Price	Ir Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale Price	Ir Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures


Related Product Guide:

Being supported by an innovative and experienced IT team, we could present technical support on pre-sales & after-sales service for Wholesale Price Ir Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran , The product will supply to all over the world, such as: Angola, Ethiopia, Greece, We hope to have long-term cooperation relationships with our clients. If you are interested in any of our products, please do not hesitate to send enquiry to us/company name. We ensure that you can be totally satisfied with our best solutions!
It can be said that this is a best producer we encountered in China in this industry, we feel lucky to work with so excellent manufacturer.
5 Stars By Klemen Hrovat from Estonia - 2017.02.28 14:19
We are old friends, the company's product quality has been always very good and this time the price is also very cheap.
5 Stars By Abigail from Zimbabwe - 2017.08.28 16:02