Wholesale Ice Point Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran

Short Description:


Product Detail

Product Tags

Related Video

Feedback (2)

Assume full responsibility to meet all demands of our clients; achieve continuous advancements by promoting the growth of our clients; become the final permanent cooperative partner of clients and maximize the interests of clients for Bimetal Temperature Calibration, Liquid Thermostatic Calibration Bath, Thermocouple Cold Junction Reference Unit, We welcome all of the clients and friends to contact us for mutual benefits. Hope to do further business with you.
Wholesale Ice Point Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran Detail:

Standard Platinum Resistance Thermometer

I. Description

The Standard Platinum Resistance Thermometer is used for compensation in standard temperature range of 13.8033k—961.8°C, and used as a standard when testing a variety of standard thermometers and high-precision thermometers. Within the above the temperature zone, it is also used directly for measuring the temperature of high-accuracy.

The Standard Platinum Resistance Thermometer measures the temperature according to the changeable regularity of the resistance temperature of the platinum.

In accordance with the regulations of ITS90, the T90 is defined by the platinum thermometer when triple point (13.8033K) of the nitrogen balance reaches to the temperature range of silver freezing point. It is indexed by using the group of required defined freezing point and reference function as well as the deviation function of the temperature interpolation.

The above temperature zoning is divided into several and decided to work normally within the sub-temperature zone by various different forms of structure of thermometers.

See the detailed thermometers in the table below:

Type

Classification

Suitable Temperature Zone

Working Length

(mm)

Temperature

WZPB-1

I

0~419.527

470±10

Medium

WZPB-1

I

83.8058K~419.527

470±10

Full

WZPB-2

II

0~419.527

470±10

Medium

WZPB-2

II

83.8058K~419.527

470±10

Full

WZPB-7

I

0~660.323

510±10

Medium

WZPB-7

I

83.8058K~660.323

510±10

Full

WZPB-8

II

0~660.323

510±10

Medium

WZPB-8

II

83.8058K~660.323

510±10

Full

Note: The Rtp of the above thermometers is 25±1.0Ω.The external diameter of quartz tubes is φ7±0.6mm.our factory also manufactures the platinum thermometer with the temperature zone of 83.8058K~660.323 as the working basic standard instrument.

II. Use Information

1. Before using, firstly, check the thermometer number to be consistent with the test certificate.

2. When using, according to the lug logo of the thermometer wire terminal, connect the wire correctly. The lug① of the red wire is connected to the current positive terminal; the lug of the yellow wire, to the current negative terminal; and the lug of the black wire, to the potential positive terminal; the lug of the green wire, to the potential negative terminal.

The following is the outline of the thermometer:

SPRT basic structure.png

3. The current should be 1MA according to the measuring of the temperature component of the thermometer.

4. For matching the electrical measurement device of the thermometer for measuring temperature, the low resistance potentiometer of grade 1 and the standard coil resistance of grade 0.1 or the measuring precise temperature bridge as well as the accessories should be used. The complete set of electrical measuring device should be guaranteed to have the sensitivity to distinguish the changing of one ten-thousandth Ohm.

5. During the process of using, preservation and transportation, try to avoid the severe mechanical vibration of the thermometer.

6. When using the first grade Standard Platinum Resistance Thermometer to test the temperature of the second grade Standard Platinum Resistance Thermometer, should follow the verification procedures approved by National Measurement Bureau.

7. The regular test of the thermometer should be done stringently according to the relevant verification procedures and regulations .


Product detail pictures:

Wholesale	Ice Point Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale	Ice Point Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale	Ice Point Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale	Ice Point Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures

Wholesale	Ice Point Thermometer Calibration	- Standard Platinum Resistance Thermometer – Panran detail pictures


Related Product Guide:

Sticking to your belief of Creating solutions of high quality and generating buddies with people from all around the world, we always put the fascination of customers to start with for Wholesale Ice Point Thermometer Calibration - Standard Platinum Resistance Thermometer – Panran , The product will supply to all over the world, such as: Amman, Karachi, Ethiopia, we always keep our credit and mutual benefit to our client, insist our high quality service to moving our clients. always welcome the our friends and clients to come and visit our company and guid our business, if you are interested in our products, you can also submit your purchase information online, and we will contact you immediately, we keep our highly sincere cooperation and wish everything in your side are all well.
We feel easy to cooperate with this company, the supplier is very responsible, thanks.There will be more in-depth cooperation.
5 Stars By April from Burundi - 2017.01.28 18:53
High Quality, High Efficiency, Creative and Integrity, worth having long-term cooperation! Looking forward to the future cooperation!
5 Stars By lucia from Bolivia - 2018.12.14 15:26